Á¦ÀÛ»ç¹Ù·Î°¡±â Ital Structures Linseis Mercury Instruments Sensovation
AULA-254.pdf Size : 2.45 MB
AULA-254´Â, ¾×»ó»óÅÂÀÇ ½Ã·á³»ÀÇ ¼öÀº³óµµ¸¦ ÃøÁ¤Çϴµ¥ »ç¿ëµÈ´Ù.
- ÀÚµ¿½Ã·áÀüó¸® ÀåÄ¡ »ç¿ë
- ±Ý ¾Æ¸»°¨¹ýÀ» »ç¿ëÇÑ ±Ø¹Ì·® ¼öÀºÃøÁ¤°¡´É(1ng/L)
- 1È¸ÃøÁ¤½Ã ÃÖ´ë 53°³ ½Ã·á ÀåÂø
- Àü±â´É PCÁ¦¾î
- ½¬¿î ¿î¿ë¹æ¹ý
Measuring principle:
UV-Absorption, cold vapor technique (CVAAS)
Analytical wavelength:
253,7 nm
UV source:
Electrodeless low-pressure mercury discharge lamp
Stabilization method:
Reference beam technique
Optical cell:
Entirely made of fused silica, ca. 230 mm length
Optical cell heating:
Approx. 50° C
Stripping gas:
argon (optional nitrogen), 4-6 l/h,stabilized with electronic mass flow controller
Autosampler:
53 place random access, turntable-type
Sample vials:
10 ml, glass; aluminium foil disc covers as accessory
Heating coil temperature:
ca. 98° C
Detection limit:
< 80 pg Hg
Measuring range:
10 ng/l ... 10 µg/l
Zero drift:
Auto zero before each measurement
Measuring duration:
Typical 60 Ð 280 seconds
Software:
AULA-WIN, WindowsTM based
Electrical power supply:
115 V / 230 V~; 50 - 60 Hz, consumption ca. 100 W (AULA 254-ASD: 120 W)
Dimensions:
Autosampler-reaction unit: 345 x 295 x 330 mm (WxHxD) Photometer: 450 x 150 x 350 mm (WxHxD)
Bench space requirements:
ca. 90 x 70 cm (W x D), excl. PC
Weight:
Approx. 14 kg total (excl. PC)
- ±Ý¾Æ¸»°¨ ÀåÂø½Ã
Preconcentration principle:
Amalgamation on gold surface and subsequent release by quick heating
1 ng/l ~ 5 µg/l
< 8 pg Hg