X-ray
generator (PC controlled) |
Output stability |
<0.01% (for 10% power supply fluctuation) |
| Max. output current |
60
mA |
| Voltage step width |
0.1
kV |
| Current step width |
0.1
mA |
| PC control serial mode |
X-ray
ON/OFF, settings of kV and mA, shutters |
| Safety devices |
Door
interlock mechaninsm, emergency push button X-ray ON lamp |
| Weight |
45
kg |
| Size |
Width
48.3 cm, height 22 cm, depth 63
cm | | |
|
X-ray
tube |
Type |
Glass,
Cu anode, fine focus |
| Focus |
0.4 x 8
mm |
| |
|
Goniometer |
Type: GD 2000 |
Independent theta/2 theta, vertical or horizontal type. |
| Variable
scanning radius |
175 -
220 mm |
| Horizontal scanning angular range |
-110°
< 2 theta < +165° |
| Smallest
selectable stepsize |
0.001°
(2 theta) |
| Scanning
speed |
0.05° -
50°/min |
| Slewing
speed |
1000°/min |
| Modes of operation |
Continuous scan, step scan, theta or 2 theta scan, fast scan, theta axis
oscillation |
| Sample
holder |
20 x 15
x 2 mm |
| Anti-divergence slits |
4°; 2°;
1°; 1/2°; 1/4° |
| Receiving slits |
0.3;
0.2; 0.1 mm |
| Soller
slits |
2°;
5° | | |
|
Detector |
Type |
Scintillation Counter NaI |
| HV/PHA |
High
voltage supply 600 - 2000 V, gain, low, central and high level
control | | |
|
Case |
Dimensions |
Width
1190mm, height 1680mm, depth 800mm |
| Leakage X-rays |
< 1
mSv/Year | | |
|