Multiplexer Module |
|
Number of sampling points: |
2 ... 24 |
Purging of sample lines: |
continuously, ca. 40 ... 80 l/h for each channel |
Flow control: |
Flow meters with alarm sensors for low flow |
Heating of valve unit: |
approx. 70 °C |
Sample line connectors: |
fittings for 4 mm i.d./ 6 mm o.d. tubing |
Valves: |
minimum dead volume |
Measuring duration: |
45 sec ... 99 minutes, adjustable separately for each channel |
Detector |
|
Measuring principle: |
UV absorption at 253,7 nm |
Type for standard range: |
Mercury-Vapor-Monitor VM-3000 |
Measuring range
(standard): |
0.1 to 100 µg/m³; 1 to 1000 µg/m³; 1 to 2000 µg/m³ |
Type for low range: |
Mercury Ultratracer UT-3000 |
Measuring range (low range): |
0.01 to 50 µg/m³ |
Sample flow: |
ca. 80 l/h |
PC connection: |
serial protocol RS 232; connection by cable (optionally by fibre optic cable);
modbus RTU; ethernet |
Mechanical |
|
Cabinet: |
metal, protection standard IP 55 (NEMA 4X, others as an option) |
Dimensions: |
from 553 x 600 x 500 mm (WxHxD) to 800 x 2000 x 600 mm; according to channel number and optional equipment (air conditioning) |
Weight: |
approx. 35 - 175 kg; depending on number of channels and optional equipment |
Software |
|
System requirement: |
PC with Windows XP, Vista |
Display of measuring data: |
actual readings numerically, filed data as time-concentration graphs |
 |
As a leading supplier of high precision analytical equipment, we strive at all times to offer top quality solutions. Our products are manufactured according to the ISO 9001 quality regulations. |
|
|