Measuring principle: |
Atomic Absorption spectrometry [AAS] |
Wavelength: |
253,7 nm |
UV source: |
Electrodeless low-pressure mercury discharge lamp |
Stabilization method: |
reference detector and thermal control |
Optical cell: |
Fused silica (Suprasil), ca. 230 mm length |
Measuring range: |
0...100 µg/m³
0...1000 ug/m³
0...2000 ug/m³ |
Sensitivity: |
0.1 µg/m³ [0.01 ppb) |
Response time (99% signal): |
1 second |
Computation of mean value: |
Automatically via three freely selectable time intervals |
Alarm: |
status contacts and optical, when concentration is exceeded, 3 levels are programmable |
Status alarms: |
measuring cell soiled, battery state, UV source exhausted |
Control pad: |
Waterproof membrane keypad |
Measurement Display: |
Graphical LC display with background lighting |
Signal outputs: |
4...20 mA for recorder;
RS 232 bidirectional for PC, data logger etc.;
parallel for printer |
Pump: |
membrane pump, approx. 1,5 L / min. |
Filter: |
PTFE, 1 µ, 47-50 mm diameter |
Power Supply: |
built-in rechargeable batteries, external 12 V DC sources, 110 ... 240 V/
(50/60 Hz) with included power supply |
Dimensions: |
425 x 195 x 305 mm (W x H x D) |
Weight: |
approx. 5.5 kg |